Software Installation
Note that installation of the GC/MS Acquisition
Software is a two-step process:
Installation of the base MassHunter GC/MS Acquisition Software B.07.00 followed
by the installation of the associated SP1 Service Pack.
Please refer to the Software Installation Instructions
(G1701_90112_GCMS_Software_Installation.pdf) in the ..\Documentation
directory on the distribution DVD for more information on how to install the
software for your particular system.
MassHunter GC/MS Acquisition Software B.07.00 SP1
Here is what is new or
changed in this revision:
New B.07.00 SP1 Acquisition
Fixes
|
MassHunter B.07.00 SP1 - SRB > LSCA
Software > GC-MS Software > GC/MS Acquisition |
||
|
KPR No. |
Issue Id |
SSB Problem Description Title |
|
309 |
200387 |
SQ: Scan plot in manual tune shows
incorrect information |
|
311 |
200066 |
SQ: Function: Foreline
Pressure; kPa ; significant figure is wrong. |
|
336 |
186999 |
SQ: Allows Dual injection for
Headspace |
|
338 |
203481 |
SQ: Selecting a different tune
file...but the temperatures do not automatically
change. |
|
363 |
205023 |
SQ: In the MSD Tune Report, Profile
and Scan windows have poor scale |
|
427 |
205177 |
SQ: In Run Method Panel, estimated
disk space is reported for D: and value is actually C: capacity |
|
437 |
205367 |
SQ: Non-responsive UI when stopping
scans, profiles, captures and ramps in Manual Tune |
|
438 |
205366 |
SQ: Changes to Profile scan speed
not persisting in Manual Tune |
|
439 |
205362 |
SQ: In Manual Tune, the Step Size Edit
box on the Profile Tab erroneously controls the step size for Scans, not
Profiles. |
|
446 |
205419 |
SQ: Strange Hivac
gauge reading when conducting commands on CI systems |
|
452 |
205472 |
SQ: During Tune, got "MS
Communication Errors" flashing on status line |
|
455 |
205477 |
SQ: AutoTune
does not use a secure default mass profiles for the tune, it uses the user
entered values from profiles |
|
458 |
205762 |
SQ: User needs to make sure Graphic
windows have stopped acquiring data before switching views |
|
470 |
206214 |
SQ: CI Reagent Gas B and Load
Method yields undefined symbol : line |
|
479 |
205871 |
SQ: .Net Exception When Clicking in
the SIM Time Segments Table |
|
480 |
205767 |
SQ: Tune Error on bfb_autotune, extra dialog saying "Starting Autotune", but nothing happens |
|
481 |
205931 |
SQ: DrugQuant:
Canceling the Save Batch Panel Causes Undefined symbol when Selecting
Simulate Batch |
|
485 |
206230 |
SQ: Sequence hangs waiting for arm
signal - intermittent |
|
486 |
207179 |
SQ: SIM/SCAN: data collected from
first SIM group only |
|
487 |
207212 |
SQ: Depending upon when override
solvent delay is selected, either a slow response or lockup could occur. |
GC and MS Firmware Updates
·
The GC firmware version must be the appropriate version
listed above or later. The GC
firmware is included on the MassHunter GC/MS Acquisition Software Installation
DVD.
1. 7890A GC firmware revision A.01.15
2. 7820 GC firmware revision A.01.12.005
3. 7810 GC firmware revision A.03.04.006
4. 7890B GC firmware revision B.02.01
·
The MS firmware is included as part of the MassHunter
GC/MS Acquisition Software installation, and is loadable via the Instrument
Configuration program.
1. 5975 series Single Quad GC/MS MS firmware revision 7.02.19
2. 5977 series Single Quad GC/MS MS firmware revision 6.00.18
3. 7000 series Triple Quad GC/MS MS firmware revision G.7000.038.
4. 7200 series Q-TOF GC/MS MS firmware revision G.7200.01.13.
·
Information on both current and previous MassHunter
GC/MS Acquisition Software firmware revisions is contained on the Installation
CD:
X:\CurrentFirmwareRevisions-en.htm
X:\PreviousReadmeMS-en.htm
Information on previous MassHunter GC/MS Acquisition Software revision is
contained on the Installation CD:
X:\PreviousReadmeMS-en.htm
For the latest PC and/or Printer
drivers, visit the HP website at http://welcome.hp.com/country/us/en/support.html
Known issues and fixed defects are listed in the Software Status Bulletin.
MassHunter GC/MS Acquisition Software B.07.00
This major revision adds
supports for the Agilent 5975 and 5977.
Here is what is new or
changed in this revision:
New Hardware/Firmware
Features
·
Support for 5975 series GC/MS; 5975A, 5975B, 5975C,
5975C TAD, 5975E, 5975T
·
Support for 5977 series GC/MS; 5977A, 5977E
·
Support for 7890B
·
Support for 7820A
·
Support for 7810A (part of the 5975T)
·
Support for the Headspace instruments (7697A and
G1888)
·
Support for 7667A Mini Thermal Desorber
New B.07.00 Acquisition
Enhancements
In addition to changes from previous versions of Acquisition, B.07.00 includes
·
New Gather Diagnostic Files… menu item is available under the
Help menu. This will create a file containing
diagnostic information that can be sent to Agilent support.
·
New Problem Steps Recorder menu item is available under the Help menu. This will run Microsoft Problem Steps Recorder (PSR.exe)
that captures mouse actions on the desktop and saves them in the recorder.zip
file.
·
Support for classic MSD ChemStation Data Analysis operation (and associated
workflows) for Single Quad Data Analysis only
1.
New Use MSD ChemStation
Data Analysis menu item is visible under the Method menu
2.
If both classic MSD ChemStation Data Analysis and MassHunter Data Analysis are
installed the menu item is visible
3.
By default, Use MSD ChemStation Data Analysis menu item is unchecked and, in
such case, MassHunter data analysis software will be used for the data analysis
portion of Run Method
·
Support for Drug Analysis via
MassHunter Quant (Sequence Table)
·
QuickLoad Methods and user-selectable Startup Method Activation
options
·
Support for the new Headspace driver
·
Parts Finder integration and desktop
icon
·
Mini TD Support
·
GC ‘Smart Technologies’
(EMF, Sleep/Wake, Fast Vent)
Operating Systems
·
Chinese Windows
7 (64) Professional SP1, Windows 7 (64) Professional
·
English Windows
7 (64) Professional SP1, Windows 7 (64) Professional
·
Japanese Windows
7 (64) Professional SP1, Windows 7 (64) Professional
Windows Vista Business SP1 is no longer supported. GC/MS
Acquisition B.05.02 is the last version that supports Windows Vista Business
SP1.
No other languages or versions of Windows are supported.
MassHunter software requires the Microsoft .NET 4.0 Framework to be
present. See
X:\Supplemental\tools\Framework\ReadMe.txt for more information on installing
it if necessary.
PC
Recommendations
·
Recommended Memory: 8 GB for Windows 7 Pro SP1 64-bit
GC and MS Firmware
·
The GC firmware version must be the appropriate
version listed below or later. The
GC firmware is included on the MassHunter GC/MS Acquisition Software
Installation DVD.
1. 7890A GC firmware revision A.01.15
2. 7820 GC firmware revision A.01.12.004
3. 7810 GC firmware revision A.03.04.005
4. 7890B GC firmware revision B.02.00
·
The MS firmware is included as part of the MassHunter
GC/MS Acquisition Software installation, and is loadable via the Instrument
Configuration program.
1. 5975 series Single Quad GC/MS MS firmware revision 7.02.18
2. 5977 series Single Quad GC/MS MS firmware revision 6.00.16
3. 7000 series Triple Quad GC/MS MS firmware revision G.7000.038 (FW 37F is
included with the B.07.00 release, but FW 38 is now available and recommended)
4. 7200 series Q-TOF GC/MS MS firmware revision G.7200.01.09.
·
More information on the current MassHunter GC/MS
Acquisition Software Add-Ons and firmware revisions
is contained on the Installation CD:
GC, MS, Autosampler Firmware Revisions
Instruments Supported
·
7200 series Q-TOF GC/MS
·
7000 series Triple Quadrupole
GC/MS
·
5975 series Single Quadrupole
GC/MS
·
5977 series Single Quadrupole
GC/MS
·
7890A GC, 7820 GC, 7810 GC (5975T)
·
7890B GC
·
7650 Autosamplers
·
7683 Autosamplers
·
7693 Autosamplers
·
G650x PAL Autosamplers (with
G9257AA Agilent CTC Control Software, B.01.03 or higher)
·
7697A and G1888 Headspace (with G7319AA Agilent Headspace
Control Software, B.01.04 or higher)
Instruments Not Supported
·
All other GCs and Mass Spec instruments
Software Supported
·
G9257AA Agilent Data System Addon
for CTC Control (rev B.01.03 or higher)
·
G7319AA Agilent Data System Addon
for Headspace Control (rev B.01.04 or higher)
·
G4600AA/G4800AA Agilent Instrument Utility Software
(rev B.01.08 or higher)
Printers and Drivers
·
LaserJet P3015d:
This printer is fully tested and is supported.
The supported driver is HP Universal Print Driver for Windows PCL6.
·
Canon LBP 3310:
Canon LBP 6700:
For Japanese Windows, these printer drivers are fully tested and supported
and are included on the Installation CD.
For the latest drivers, visit the HP website at http://welcome.hp.com/country/us/en/support.html
Information on previous MassHunter GC/MS Acquisition Software revision is
contained on the Installation CD:
Previous Software Revision ReadMeMS file
For the latest PC and/or Printer
drivers, visit the HP website at http://welcome.hp.com/country/us/en/support.html
Known issues and fixed defects are listed in the Software Status Bulletin.